本文整理汇总了Python中tube.calibrate函数的典型用法代码示例。如果您正苦于以下问题:Python calibrate函数的具体用法?Python calibrate怎么用?Python calibrate使用的例子?那么恭喜您, 这里精选的函数代码示例或许可以为您提供帮助。
在下文中一共展示了calibrate函数的15个代码示例,这些例子默认根据受欢迎程度排序。您可以为喜欢或者感觉有用的代码点赞,您的评价将有助于系统推荐出更棒的Python代码示例。
示例1: CalibrateWish
def CalibrateWish(RunNumber, PanelNumber):
'''
:param RunNumber: is the run number of the calibration.
:param PanelNumber: is a string of two-digit number of the panel being calibrated
'''
# == Set parameters for calibration ==
previousDefaultInstrument = mantid.config['default.instrument']
mantid.config['default.instrument'] = "WISH"
filename = str(RunNumber)
CalibratedComponent = 'WISH/panel' + PanelNumber
# Get calibration raw file and integrate it
print("Loading", filename)
rawCalibInstWS = mantid.Load(filename) # 'raw' in 'rawCalibInstWS' means unintegrated.
CalibInstWS = mantid.Integration(rawCalibInstWS, RangeLower=1, RangeUpper=20000)
mantid.DeleteWorkspace(rawCalibInstWS)
print("Created workspace (CalibInstWS) with integrated data from run and instrument to calibrate")
# Give y-positions of slit points (gotten for converting first tube's slit point to Y)
# WISH instrument has a particularity. It is composed by a group of upper tubes and lower tubes,
# they are disposed 3 milimiters in difference one among the other
lower_tube = numpy.array([-0.41, -0.31, -0.21, -0.11, -0.02, 0.09, 0.18, 0.28, 0.39])
upper_tube = numpy.array(lower_tube + 0.003)
funcForm = 9 * [1] # 9 gaussian peaks
print("Created objects needed for calibration.")
# Get the calibration and put it into the calibration table
# calibrate the lower tubes
calibrationTable, peakTable = tube.calibrate(CalibInstWS, CalibratedComponent, lower_tube, funcForm,
rangeList=list(range(0, 76)), outputPeak=True)
# calibrate the upper tubes
calibrationTable, peakTable = tube.calibrate(CalibInstWS, CalibratedComponent, upper_tube, funcForm,
rangeList=list(range(76, 152)),
calibTable=calibrationTable,
# give the calibration table to append data
outputPeak=peakTable # give peak table to append data
)
print("Got calibration (new positions of detectors)")
# Apply the calibration
mantid.ApplyCalibration(Workspace=CalibInstWS, PositionTable=calibrationTable)
print("Applied calibration")
# == Save workspace ==
# uncomment these lines to save the workspace
# nexusName = "TubeCalibDemoWish" + PanelNumber + "Result.nxs"
# mantid.SaveNexusProcessed(CalibInstWS, 'TubeCalibDemoWishResult.nxs', "Result of Running TubeCalibWishMerlin_Simple.py")
# print("saved calibrated workspace (CalibInstWS) into Nexus file", nexusName)
# == Reset dafault instrument ==
mantid.config['default.instrument'] = previousDefaultInstrument
示例2: provideTheExpectedValue
def provideTheExpectedValue(filename):
"""
Giving the expected value for the position of the peaks in pixel.
The :func:`~Examples.minimalInput` let to the calibrate to guess the position of the pixels
among the tubes. Altough it works nicelly, providing these expected values may improve the results.
This is done through the **fitPar** parameter.
"""
from tube_calib_fit_params import TubeCalibFitParams
CalibInstWS = loadingStep(filename)
# == Set parameters for calibration ==
# Set what we want to calibrate (e.g whole intrument or one door )
CalibratedComponent = 'MAPS' # Calibrate all
# define the known positions and function factor (edge, peak, peak, peak, edge)
knownPos, funcFactor = [-0.50,-0.16,-0.00, 0.16, 0.50 ],[2,1,1,1,2]
# the expected positions in pixels for the special points
expectedPositions = [4.0, 85.0, 128.0, 161.0, 252.0]
fitPar = TubeCalibFitParams(expectedPositions)
fitPar.setAutomatic(True)
# == Get the calibration and put results into calibration table ==
calibrationTable = tube.calibrate(CalibInstWS, CalibratedComponent, knownPos, funcFactor,
fitPar=fitPar)
# == Apply the Calibation ==
ApplyCalibration( Workspace=CalibInstWS, PositionTable=calibrationTable)
示例3: minimalInput
def minimalInput(filename):
"""
Simplest way of calling :func:`tube.calibrate`
The minimal input for the calibration is the integrated workspace
and the knwon positions.
Eventhough it is easy to call, the calibration performs well, but there are ways to improve
the results, as it is explored after.
.. image:: /images/outputOfMinimalInput.png
"""
CalibInstWS = loadingStep(filename)
# == Set parameters for calibration ==
# Set what we want to calibrate (e.g whole intrument or one door )
CalibratedComponent = 'MAPS' # Calibrate all
# define the known positions and function factor (edge, peak, peak, peak, edge)
knownPos, funcFactor = [-0.50,-0.16,-0.00, 0.16, 0.50 ],[2,1,1,1,2]
# == Get the calibration and put results into calibration table ==
calibrationTable = tube.calibrate(CalibInstWS, CalibratedComponent, knownPos, funcFactor)
# == Apply the Calibation ==
ApplyCalibration( Workspace=CalibInstWS, PositionTable=calibrationTable)
示例4: improvingCalibrationOfListOfTubes
def improvingCalibrationOfListOfTubes(filename):
"""
Analysing the result of provideTheExpectedValue it was seen that the calibration
of some tubes was not good.
.. note::
This method list some of them, there are a group belonging to window B2 that shows
only 2 peaks that are not dealt with here.
If first plot the bad ones using the **plotTube** option. It them, find where they fail, and how
to correct their peaks, using the **overridePeaks**.
If finally, applies the calibration again with the points corrected.
"""
from tube_calib_fit_params import TubeCalibFitParams
not_good = [19,37, 71, 75, 181, 186, 234, 235, 245, 273, 345]
CalibInstWS = loadingStep(filename)
# == Set parameters for calibration ==
# Set what we want to calibrate (e.g whole intrument or one door )
CalibratedComponent = 'MAPS' # Calibrate all
# define the known positions and function factor (edge, peak, peak, peak, edge)
knownPos, funcFactor = [-0.50,-0.16,-0.00, 0.16, 0.50 ],[2,1,1,1,2]
# the expected positions in pixels for the special points
expectedPositions = [4.0, 85.0, 128.0, 161.0, 252.0]
fitPar = TubeCalibFitParams(expectedPositions)
fitPar.setAutomatic(True)
# == Get the calibration and put results into calibration table ==
#calibrationTable, peakTable= tube.calibrate(CalibInstWS, CalibratedComponent, knownPos, funcFactor,
# fitPar=fitPar, outputPeak=True, plotTube=not_good, rangeList=not_good)
#CalibInstWS = loadingStep(filename)
# it is defined as the mean values around the neighbours
define_peaks = {19:[10, 80.9771, 123.221, 164.993, 245.717], # the first one was bad
37: [6.36, 80.9347, 122.941, 165.104, 248.32], # the first one was bad
71: [8.62752, 85.074, 124.919, 164.116, 246.82 ], # the last one was bad - check if we can inprove
75: [14.4285, 90.087, 128.987, 167.047, 242.62], # the last one was bad - check if we can inprove
181: [11.726, 94.0496, 137.816, 180, 255], # the third peak was lost
186:[11.9382, 71.5203, 107, 147.727, 239.041], #lost the second peak
234: [4.84, 82.7824, 123.125, 163.945, 241.877], # the first one was bad
235: [4.84, 80.0077, 121.002, 161.098, 238.502], # the first one was bad
245: [9.88089, 93.0593, 136.911, 179.5, 255], # the third peak was bad
273: [18.3711, 105.5, 145.5, 181.6, 243.252], # lost first and third peaks
345: [4.6084, 87.0351, 128.125, 169.923, 245.3] # the last one was bad
}
calibrationTable, peakTable= tube.calibrate(CalibInstWS, CalibratedComponent, knownPos, funcFactor,
fitPar=fitPar, outputPeak=True, overridePeaks=define_peaks)
ApplyCalibration( Workspace=CalibInstWS, PositionTable=calibrationTable)
示例5: changeMarginAndExpectedValue
def changeMarginAndExpectedValue(filename):
"""
To fit correcly, it is important to have a good window around the peak. This windown is defined
by the **margin** parameter.
This examples shows how the results worsen if we change the margin from its default value **15**
to **10**.
It shows how to see the fitted values using the **plotTube** parameter.
It will also output the peaks position and save them, through the **outputPeak** option and
the :func:`tube.savePeak` method.
An example of the fitted data compared to the acquired data to find the peaks positions:
.. image:: /images/calibratePlotFittedData.png
The result deteriorate, as you can see:
.. image:: /images/calibrateChangeMarginAndExpectedValue.png
"""
from tube_calib_fit_params import TubeCalibFitParams
CalibInstWS = loadingStep(filename)
# == Set parameters for calibration ==
# Set what we want to calibrate (e.g whole intrument or one door )
CalibratedComponent = 'MAPS' # Calibrate all
# define the known positions and function factor (edge, peak, peak, peak, edge)
knownPos, funcFactor = [-0.50,-0.16,-0.00, 0.16, 0.50 ],[2,1,1,1,2]
# the expected positions in pixels for the special points
expectedPositions = [4.0, 85.0, 128.0, 161.0, 252.0]
fitPar = TubeCalibFitParams(expectedPositions)
fitPar.setAutomatic(True)
# == Get the calibration and put results into calibration table ==
calibrationTable, peakTable= tube.calibrate(CalibInstWS, CalibratedComponent, knownPos, funcFactor,
fitPar=fitPar, plotTube=[1,10,100], outputPeak=True, margin=10)
# == Apply the Calibation ==
ApplyCalibration( Workspace=CalibInstWS, PositionTable=calibrationTable)
tube.savePeak(peakTable, 'TubeDemoMaps01.txt')
示例6: calibrateB2Window
def calibrateB2Window(filename):
"""
There are among the B2 window tubes, some tubes that are showing only 2 strips.
Those tubes must be calibrated separated, as the known positions are not valid.
This example calibrate them, using only 4 known values: 2 edges and 2 peaks.
Run this example, and them see the worksapce in the calibrated instrument and you will see
how it worked.
The picture shows the output, look that only a section of the B2 Window was calibrated.
.. image:: /images/calibrateB2Window.png
"""
from tube_calib_fit_params import TubeCalibFitParams
# b2 with 2 peaks range
b2_range = list(range(196, 212)) + list(range(222, 233))
CalibInstWS = loadingStep(filename)
# == Set parameters for calibration ==
# Set what we want to calibrate (e.g whole instrument or one door )
CalibratedComponent = 'MAPS' # Calibrate all
# define the known positions and function factor (edge, peak, peak, peak, edge)
knownPos, funcFactor = [-0.50, -0.16, 0.16, 0.50], [2, 1, 1, 2]
# the expected positions in pixels for the special points
expectedPositions = [4.0, 85.0, 161.0, 252.0]
fitPar = TubeCalibFitParams(expectedPositions)
fitPar.setAutomatic(True)
# == Get the calibration and put results into calibration table ==
calibrationTable, peakTable = tube.calibrate(CalibInstWS, CalibratedComponent, knownPos, funcFactor,
fitPar=fitPar, outputPeak=True, plotTube=[b2_range[0], b2_range[-1]],
rangeList=b2_range)
mantid.ApplyCalibration(Workspace=CalibInstWS, PositionTable=calibrationTable)
示例7: workspace
print "Created workspace (CalibInstWS) with integrated data from run and instrument to calibrate"
# == Create Objects needed for calibration ==
# Specify components to calibrate
thisTubeSet = TubeSpec(CalibInstWS)
thisTubeSet.setTubeSpecByString(CalibratedComponent1)
thisTubeSet.setTubeSpecByString(CalibratedComponent2)
# Specify the known positions
knownPos = [-0.50,-0.16,-0.00, 0.16, 0.50 ]
funcForm = [2,1,1,1,2]
print "Created objects needed for calibration."
# == Get the calibration and put results into calibration table ==
calibrationTable, peakTable = tube.calibrate(CalibInstWS, thisTubeSet, knownPos, funcForm,
outputPeak=True)
print "Got calibration (new positions of detectors) "
# == Apply the Calibation ==
ApplyCalibration( Workspace=CalibInstWS, PositionTable=calibrationTable)
print "Applied calibration"
# == Save workspace ==
#SaveNexusProcessed( CalibInstWS, path+'TubeCalibDemoMapsResult.nxs',"Result of Running TCDemoMaps.py")
print "saved calibrated workspace (CalibInstWS) into Nexus file TubeCalibDemoMapsResult.nxs"
示例8: completeCalibration
def completeCalibration(filename):
"""
This example shows how to use some properties of calibrate method to
join together the calibration done in :func:`provideTheExpectedValue`,
and improved in :func:`calibrateB2Window`, and :func:`improvingCalibrationOfListOfTubes`.
It also improves the result of the calibration because it deals with the E door. The
aquired data cannot be used to calibrate the E door, and trying to do so, produces a bad
result. In this example, the tubes inside the E door are excluded to the calibration.
Using the '''rangeList''' option.
"""
# first step, load the workspace
from tube_calib_fit_params import TubeCalibFitParams
CalibInstWS = loadingStep(filename)
# == Set parameters for calibration ==
# Set what we want to calibrate (e.g whole intrument or one door )
CalibratedComponent = 'MAPS' # Calibrate all
# define the known positions and function factor (edge, peak, peak, peak, edge)
knownPos, funcFactor = [-0.50,-0.16,-0.00, 0.16, 0.50 ],[2,1,1,1,2]
# the expected positions in pixels for the special points
expectedPositions = [4.0, 85.0, 128.0, 161.0, 252.0]
fitPar = TubeCalibFitParams(expectedPositions)
fitPar.setAutomatic(True)
#execute the improvingCalibrationOfListOfTubes excluding the range of b2 window
# correct the definition of the peaks for the folowing indexes
#define_peaks = {19:[10, 80.9771, 123.221, 164.993, 245.717], # the first one was bad
# 37: [6.36, 80.9347, 122.941, 165.104, 248.32], # the first one was bad
# 71: [8.62752, 85.074, 124.919, 164.116, 246.82 ], # the last one was bad - check if we can inprove
# 75: [14.4285, 90.087, 128.987, 167.047, 242.62], # the last one was bad - check if we can inprove
# 181: [11.726, 94.0496, 137.816, 180, 255], # the third peak was lost
# 186:[11.9382, 71.5203, 107, 147.727, 239.041], #lost the second peak
# 234: [4.84, 82.7824, 123.125, 163.945, 241.877], # the first one was bad
# 235: [4.84, 80.0077, 121.002, 161.098, 238.502], # the first one was bad
# 245: [9.88089, 93.0593, 136.911, 179.5, 255], # the third peak was bad
# 273: [18.3711, 105.5, 145.5, 181.6, 243.252],# lost first and third peaks
# 345: [4.6084, 87.0351, 128.125, 169.923, 245.3]} # the last one was bad
define_peaks = {19:[10, 80.9771, 123.221, 164.993, 245.717],\
37: [6.36, 80.9347, 122.941, 165.104, 248.32],\
71: [8.62752, 85.074, 124.919, 164.116, 246.82 ],\
75: [14.4285, 90.087, 128.987, 167.047, 242.62],\
181: [11.726, 94.0496, 137.816, 180, 255],\
186:[11.9382, 71.5203, 107, 147.727, 239.041],\
234: [4.84, 82.7824, 123.125, 163.945, 241.877],\
235: [4.84, 80.0077, 121.002, 161.098, 238.502],\
245: [9.88089, 93.0593, 136.911, 179.5, 255],\
273: [18.3711, 105.5, 145.5, 181.6, 243.252],\
345: [4.6084, 87.0351, 128.125, 169.923, 245.3]}
b2_window = range(196,212) + range(222,233)
complete_range = range(648)
# this data can not be used to calibrate the E1 window, so, let's remove it.
e1_window = range(560,577)
aux = numpy.setdiff1d(complete_range, b2_window)
# the group that have 3 stripts are all the tubes except the b2 window and e window.
range_3_strips = numpy.setdiff1d(aux, e1_window)
calibrationTable, peak3Table= tube.calibrate(CalibInstWS, CalibratedComponent, knownPos, funcFactor,\
fitPar=fitPar, outputPeak=True, overridePeaks=define_peaks, rangeList=range_3_strips)
# now calibrate the b2_window REMOVE SECOND PEAK
# define the known positions and function factor (edge, peak, peak, edge)
knownPos, funcFactor = [-0.50,-0.16, 0.16, 0.50 ],[2,1,1,2]
# the expected positions in pixels for the special points
expectedPositions = [4.0, 85.0, 161.0, 252.0]
fitPar = TubeCalibFitParams(expectedPositions)
fitPar.setAutomatic(True)
# apply the calibration for the b2_window 2 strips values
calibrationTable, peak2Table = tube.calibrate(CalibInstWS, CalibratedComponent,
knownPos, #these parameters now have only 4 points
funcFactor,
fitPar=fitPar,
outputPeak=True,
calibTable = calibrationTable, # it will append to the calibTable
rangeList = b2_window)
ApplyCalibration( Workspace=CalibInstWS, PositionTable=calibrationTable)
示例9: findThoseTubesThatNeedSpecialCareForCalibration
def findThoseTubesThatNeedSpecialCareForCalibration(filename):
"""
The example :func:`provideTheExpectedValue` has shown its capability to calibrate almost
all tubes, but, as explored in the :func:`improvingCalibrationOfListOfTubes` and
:func:`improvingCalibrationSingleTube` there are
some tubes that could not be calibrated using that method.
The goal of this method is to show one way to find the tubes that will require special care.
It will first perform the same calibration seen in :func:`provideTheExpectedValue`,
them, it will process the **peakTable** output of the calibrate method when enabling the
parameter **outputPeak**.
It them creates the Peaks workspace, that is the diffence of the peaks position from the
expected values of the peaks positions for all the tubes. This allows to spot what are the
tubes whose fitting are outliers in relation to the others.
.. image:: /images/plotingPeaksDifference.png
The final result for this method is to output using **plotTube** the result of the fitting
to all the 'outliers' tubes.
"""
from tube_calib_fit_params import TubeCalibFitParams
CalibInstWS = loadingStep(filename)
# == Set parameters for calibration ==
# Set what we want to calibrate (e.g whole intrument or one door )
CalibratedComponent = 'MAPS' # Calibrate all
# define the known positions and function factor (edge, peak, peak, peak, edge)
knownPos, funcFactor = [-0.50,-0.16,-0.00, 0.16, 0.50 ],[2,1,1,1,2]
# the expected positions in pixels for the special points
expectedPositions = [4.0, 85.0, 128.0, 161.0, 252.0]
fitPar = TubeCalibFitParams(expectedPositions)
fitPar.setAutomatic(True)
# == Get the calibration and put results into calibration table ==
calibrationTable, peakTable= tube.calibrate(CalibInstWS, CalibratedComponent, knownPos, funcFactor,
fitPar=fitPar, outputPeak=True)
# == now, lets investigate the peaks
#parsing the peakTable to produce a numpy array with dimension (number_of_tubes x number_of_peaks)
print 'parsing the peak table'
n = len(peakTable)
peaksId = n*['']
data = numpy.zeros((n,5))
line = 0
for row in peakTable:
data_row = [row['Peak%d'%(i)] for i in [1,2,3,4,5]]
data[line,:] = data_row
peaksId[line] = row['TubeId']
line+=1
# data now has all the peaks positions for each tube
# the mean value is the expected value for the peak position for each tube
expected_peak_pos = numpy.mean(data,axis=0)
#calculate how far from the expected position each peak position is
distance_from_expected = numpy.abs(data - expected_peak_pos)
print 'Creating the Peaks Workspace that shows the distance from the expected value for all peaks for each tube'
# Let's see these peaks:
Peaks = CreateWorkspace(range(n),distance_from_expected,NSpec=5)
# plot all the 5 peaks for Peaks Workspace. You will see that most of the tubes differ
# at most 12 pixels from the expected values.
#so let's investigate those that differ more than 12
# return an array with the indexes for the first axis which is the tube indentification
check = numpy.where(distance_from_expected > 12)[0]
#remove repeated values
#select only those tubes inside the problematic_tubes
problematic_tubes = list(set(check))
print 'Tubes whose distance is far from the expected value: ', problematic_tubes
print 'Calibrating again only these tubes'
#let's confir that our suspect works
CalibInstWS = loadingStep(filename)
calibrationTable = tube.calibrate(CalibInstWS, CalibratedComponent, knownPos, funcFactor,\
fitPar=fitPar, rangeList= problematic_tubes, plotTube=problematic_tubes)
示例10: Load
import os
from mantid.simpleapi import *
from mantid import api
import numpy as np
import tube
ws = Load("MER31013.n003")
ws = Integration(ws,1500,9000)
pos = [0.84,0.44,0.02,-0.45,-0.65]
tt = [1,1,1,1,1]
calt = tube.calibrate(ws,"MERLIN",pos,tt)
示例11: calibrateMerlin
def calibrateMerlin(filename):
# == Set parameters for calibration ==
rangeLower = 3000 # Integrate counts in each spectra from rangeLower to rangeUpper
rangeUpper = 20000 #
# Get calibration raw file and integrate it
rawCalibInstWS = LoadRaw(filename) #'raw' in 'rawCalibInstWS' means unintegrated.
print "Integrating Workspace"
CalibInstWS = Integration( rawCalibInstWS, RangeLower=rangeLower, RangeUpper=rangeUpper )
DeleteWorkspace(rawCalibInstWS)
print "Created workspace (CalibInstWS) with integrated data from run and instrument to calibrate"
# the known positions are given in pixels inside the tubes and transformed to provide the positions
# with the center of the tube as the origin
knownPositions = 2.92713867188*(numpy.array([ 27.30074322, 92.5, 294.65178585, 362.37861919 , 512.77103043 ,663.41425323, 798.3223896, 930.9, 997.08480835])/1024 - 0.5)
funcForm = numpy.array([2,2,1,1,1,1,1,2,2],numpy.int8)
# The calibration will follow different steps for sets of tubes
# For the door9, the best points to define the known positions are the 1st edge, 5 peaks, last edge.
points7 = knownPositions[[0,2,3,4,5,6,8]]
points7func = funcForm[[0,2,3,4,5,6,8]]
door9pos = points7
door9func = points7func
CalibratedComponent = 'MERLIN/door9' # door9
# == Get the calibration and put results into calibration table ==
# also put peaks into PeakFile
calibrationTable, peakTable = tube.calibrate(CalibInstWS, CalibratedComponent, door9pos, door9func,
outputPeak=True,
margin=30,
rangeList=range(20) # because 20, 21, 22, 23 are defective detectors
)
print "Got calibration (new positions of detectors) and put slit peaks into file TubeDemoMerlin01.txt"
analisePeakTable(peakTable, 'door9_tube1_peaks')
# For the door8, the best points to define the known positions are the 1st edge, 5 peaks, last_edge
door8pos = points7
door8func = points7func
CalibratedComponent = 'MERLIN/door8'
calibrationTable, peakTable = tube.calibrate(CalibInstWS, CalibratedComponent, door8pos,
door8func,
outputPeak = True, #change to peakTable to append to peakTable
calibTable = calibrationTable,
margin = 30)
analisePeakTable(peakTable, 'door8_peaks')
# For the doors 7,6,5,4, 2, 1 we may use the 9 points
doorpos = knownPositions
doorfunc = funcForm
CalibratedComponent = ['MERLIN/door%d'%(i) for i in [7,6,5,4, 2, 1]]
calibrationTable, peakTable = tube.calibrate(CalibInstWS, CalibratedComponent, doorpos,\
doorfunc,\
outputPeak = True,\
calibTable = calibrationTable,\
margin = 30)
analisePeakTable(peakTable, 'door1to7_peaks')
# The door 3 is a special case, because it is composed by diffent kind of tubes.
# door 3 tubes: 5_8, 5_7, 5_6, 5_5, 5_4, 5_3, 5_2, 5_1, 4_8, 4_7, 4_6, 4_5, 4_4, 4_3, 4_2, 4_1, 3_8, 3_7, 3_6, 3_5, 3_4
# obeys the same rules as the doors 7, 6, 5, 4, 2, 1
# For the tubes 3_3, 3_2, 3_1 -> it is better to skip the central peak
# For the tubes 1_x (smaller tube below), it is better to take the final part of known positions: peak4,peak5,edge6,edge7
# For the tubes 2_x (smaller tube above, it is better to take the first part of known positions: edge1, edge2, peak1,peak2
# NOTE: the smaller tubes they have length = 1.22879882813, but 1024 detectors
# so we have to correct the known positiosn by multiplying by its lenght and dividing by the longer dimension
from tube_calib_fit_params import TubeCalibFitParams
# calibrating tubes 1_x
CalibratedComponent = ['MERLIN/door3/tube_1_%d'%(i) for i in range(1,9)]
half_diff_center = (2.92713867188 -1.22879882813)/2 # difference among the expected center position for both tubes
# here a little bit of attempts is necessary. The efective center position and lengh is different for the calibrated tube, that
# is the reason, the calibrated values of the smaller tube does not seems aligned with the others. By, finding the 'best' half_diff_center
# value, the alignment occurs nicely.
half_diff_center = 0.835 #
# the knownpositions were given with the center of the bigger tube as origin, to convert
# to the center of the upper tube as origin is necessary to subtract them with the half_diff_center
doorpos = knownPositions[[5,6,7,8]] - half_diff_center
doorfunc = [1,1,2,2]
# for the smal tubes, automatically searching for the peak position in pixel was not working quite well,
# so we will give the aproximate position for these tubes through fitPar argument
fitPar = TubeCalibFitParams([216, 527, 826, 989])
fitPar.setAutomatic(True)
calibrationTable, peakTable = tube.calibrate(CalibInstWS, CalibratedComponent, doorpos,\
doorfunc,\
outputPeak = True,\
fitPar = fitPar,\
calibTable = calibrationTable,\
margin = 30)
analisePeakTable(peakTable, 'door3_tube1_peaks')
# calibrating tubes 2_x
CalibratedComponent = ['MERLIN/door3/tube_2_%d'%(i) for i in range(1,9)]
# the knownpositions were given with the center of the bigger tube as origin, to convert
#.........这里部分代码省略.........
示例12: TubeCalibFitParams
# == Create Objects needed for calibration ==
# The positions of the shadows and ends here are an intelligent guess.
# First array gives positions in Metres and second array gives type 1=Gaussian peak 2=edge.
# See http://www.mantidproject.org/IdealTube for details.
knownPos = [-0.50, -0.165, -0.00, 0.165, 0.50]
funcForm = [2, 1, 1, 1, 2]
# Get fitting parameters
fitPar = TubeCalibFitParams(ExpectedPositions, ExpectedHeight, ExpectedWidth)
fitPar.setAutomatic(True)
print("Created objects needed for calibration.")
# == Get the calibration and put results into calibration table ==
# also put peaks into PeakFile
calibrationTable, peakTable = tube.calibrate(CalibInstWS, CalibratedComponent, knownPos, funcForm,
fitPar=fitPar, outputPeak=True)
print("Got calibration (new positions of detectors) ")
# == Apply the Calibation ==
mantid.ApplyCalibration(Workspace=CalibInstWS, PositionTable=calibrationTable)
print("Applied calibration")
# == Save workspace ==
mantid.SaveNexusProcessed(CalibInstWS, 'TubeCalibDemoMapsResult.nxs', "Result of Running TCDemoMaps_B1.py")
print("saved calibrated workspace (CalibInstWS) into Nexus file TubeCalibDemoMapsResult.nxs")
# == Save Peak File ==
tube.savePeak(peakTable, 'TubeDemoMaps01.txt')
示例13: print
CalibInstWS = mantid.Integration( rawCalibInstWS, RangeLower=1, RangeUpper=20000 )
mantid.DeleteWorkspace(rawCalibInstWS)
print("Created workspace (CalibInstWS) with integrated data from run and instrument to calibrate")
CalibratedComponent = 'WISH/panel03/tube038'
# Set fitting parameters
eP = [65.0, 113.0, 161.0, 209.0, 257.0, 305.0, 353.0, 401.0, 449.0]
ExpectedHeight = 2000.0 # Expected Height of Gaussian Peaks (initial value of fit parameter)
ExpectedWidth = 32.0 # Expected width of Gaussian peaks in pixels (initial value of fit parameter)
fitPar = TubeCalibFitParams( eP, ExpectedHeight, ExpectedWidth )
fitPar.setAutomatic(True)
print("Created objects needed for calibration.")
func_form = 9*[1]
# Use first tube as ideal tube
tube1 = TubeSpec(CalibInstWS)
tube1.setTubeSpecByString('WISH/panel03/tube038')
iTube = tube_calib.constructIdealTubeFromRealTube( CalibInstWS, tube1, fitPar, func_form)
known_pos = iTube.getArray()
print(known_pos)
# Get the calibration and put it into the calibration table
calibrationTable = tube.calibrate( CalibInstWS, 'WISH/panel03', known_pos, func_form, fitPar=fitPar)
print("Got calibration (new positions of detectors)")
#Apply the calibration
mantid.ApplyCalibration(Workspace=CalibInstWS, PositionTable=calibrationTable)
print("Applied calibration")
示例14: CalibrateMerlin
def CalibrateMerlin(RunNumber):
# == Set parameters for calibration ==
previousDefaultInstrument = mantid.config['default.instrument']
mantid.config['default.instrument'] = "MERLIN"
filename = str(RunNumber) # Name of calibration run.
rangeLower = 3000 # Integrate counts in each spectra from rangeLower to rangeUpper
rangeUpper = 20000 #
# Set parameters for ideal tube.
Left = 2.0 # Where the left end of tube should be in pixels (target for AP)
Centre = 512.5 # Where the centre of the tube should be in pixels (target for CP)
Right = 1023.0 # Where the right of the tube should be in pixels (target for BP)
ActiveLength = 2.9 # Active length of tube in Metres
# Set initial parameters for peak finding
ExpectedHeight = 1000.0 # Expected Height of Gaussian Peaks (initial value of fit parameter)
ExpectedWidth = 32.0 # Expected width of centre peak in Pixels (initial value of fit parameter)
ExpectedPositions = [35.0, 512.0,
989.0] # Expected positions of the edges and peak in pixels (initial values of fit parameters)
# Set what we want to calibrate (e.g whole intrument or one door )
CalibratedComponent = 'MERLIN' # Calibrate door 2
# Get calibration raw file and integrate it
print(filename)
rawCalibInstWS = mantid.LoadRaw(filename)
# 'raw' in 'rawCalibInstWS' means unintegrated.
print("Integrating Workspace")
CalibInstWS = mantid.Integration(rawCalibInstWS, RangeLower=rangeLower, RangeUpper=rangeUpper)
mantid.DeleteWorkspace(rawCalibInstWS)
print("Created workspace (CalibInstWS) with integrated data from run and instrument to calibrate")
# == Create Objects needed for calibration ==
## In the merlin case, the positions are usually given in pixels, instead of being given in
## meters, to convert to meter and put the origin in the center, we have to apply the following
## transformation:
##
## pos = pixel * length/npixels - length/2 = length (pixel/npixels - 1/2)
##
## for merlin: npixels = 1024
knownPos = ActiveLength * (numpy.array([Left, Centre, Right]) / 1024.0 - 0.5)
funcForm = 3 * [1]
# Get fitting parameters
fitPar = TubeCalibFitParams(ExpectedPositions, ExpectedHeight, ExpectedWidth, margin=40)
print("Created objects needed for calibration.")
# == Get the calibration and put results into calibration table ==
# also put peaks into PeakFile
calibrationTable, peakTable = tube.calibrate(CalibInstWS, CalibratedComponent, knownPos, funcForm,
outputPeak=True, fitPar=fitPar, plotTube=list(range(0, 280, 20)))
print("Got calibration (new positions of detectors) and put slit peaks into file TubeDemoMerlin01.txt")
# == Apply the Calibation ==
mantid.ApplyCalibration(Workspace=CalibInstWS, PositionTable=calibrationTable)
print("Applied calibration")
# == Save workspace ==
# mantid.SaveNexusProcessed(CalibInstWS, 'TubeCalibDemoMerlinResult.nxs', "Result of Running TubeCalibDemoMerlin_Simple.py")
# print("saved calibrated workspace (CalibInstWS) into Nexus file TubeCalibDemoMerlinResult.nxs")
# == Reset dafault instrument ==
mantid.config['default.instrument'] = previousDefaultInstrument
示例15: runTest
def runTest(self):
# This script calibrates WISH using known peak positions from
# neutron absorbing bands. The workspace with suffix "_calib"
# contains calibrated data. The workspace with suxxic "_corrected"
# contains calibrated data with known problematic tubes also corrected
ws = mantid.LoadNexusProcessed(Filename="WISH30541_integrated.nxs")
# This array defines the positions of peaks on the detector in
# meters from the center (0)
# For wish this is calculated as follows:
# Height of all 7 bands = 0.26m => each band is separated by 0.260 / 6 = 0.4333m
# The bands are on a cylinder diameter 0.923m. So we can work out the angle as
# (0.4333 * n) / (0.923 / 2) where n is the number of bands above (or below) the
# center band.
# Putting this together with the distance to the detector tubes (2.2m) we get
# the following: (0.4333n) / 0.4615 * 2200 = Expected peak positions
# From this we can show there should be 5 peaks (peaks 6 + 7 are too high/low)
# at: 0, 0.206, 0.413 respectively (this is symmetrical so +/-)
peak_positions = np.array([-0.413, -0.206, 0, 0.206, 0.413])
funcForm = 5 * [1] # 5 gaussian peaks
fitPar = TubeCalibFitParams([59, 161, 258, 353, 448])
fitPar.setAutomatic(True)
instrument = ws.getInstrument()
spec = TubeSpec(ws)
spec.setTubeSpecByString(instrument.getFullName())
idealTube = IdealTube()
idealTube.setArray(peak_positions)
# First calibrate all of the detectors
calibrationTable, peaks = tube.calibrate(ws, spec, peak_positions, funcForm, margin=15,
outputPeak=True, fitPar=fitPar)
self.calibration_table = calibrationTable
def findBadPeakFits(peaksTable, threshold=10):
""" Find peaks whose fit values fall outside of a given tolerance
of the mean peak centers across all tubes.
Tubes are defined as have a bad fit if the absolute difference
between the fitted peak centers for a specific tube and the
mean of the fitted peak centers for all tubes differ more than
the threshold parameter.
@param peakTable: the table containing fitted peak centers
@param threshold: the tolerance on the difference from the mean value
@return A list of expected peak positions and a list of indices of tubes
to correct
"""
n = len(peaksTable)
num_peaks = peaksTable.columnCount() - 1
column_names = ['Peak%d' % i for i in range(1, num_peaks + 1)]
data = np.zeros((n, num_peaks))
for i, row in enumerate(peaksTable):
data_row = [row[name] for name in column_names]
data[i, :] = data_row
# data now has all the peaks positions for each tube
# the mean value is the expected value for the peak position for each tube
expected_peak_pos = np.mean(data, axis=0)
# calculate how far from the expected position each peak position is
distance_from_expected = np.abs(data - expected_peak_pos)
check = np.where(distance_from_expected > threshold)[0]
problematic_tubes = list(set(check))
print("Problematic tubes are: " + str(problematic_tubes))
return expected_peak_pos, problematic_tubes
def correctMisalignedTubes(ws, calibrationTable, peaksTable, spec, idealTube, fitPar, threshold=10):
""" Correct misaligned tubes due to poor fitting results
during the first round of calibration.
Misaligned tubes are first identified according to a tolerance
applied to the absolute difference between the fitted tube
positions and the mean across all tubes.
The FindPeaks algorithm is then used to find a better fit
with the ideal tube positions as starting parameters
for the peak centers.
From the refitted peaks the positions of the detectors in the
tube are recalculated.
@param ws: the workspace to get the tube geometry from
@param calibrationTable: the calibration table output from running calibration
@param peaksTable: the table containing the fitted peak centers from calibration
@param spec: the tube spec for the instrument
@param idealTube: the ideal tube for the instrument
@param fitPar: the fitting parameters for calibration
@param threshold: tolerance defining is a peak is outside of the acceptable range
@return table of corrected detector positions
"""
table_name = calibrationTable.name() + 'Corrected'
corrections_table = mantid.CreateEmptyTableWorkspace(OutputWorkspace=table_name)
#.........这里部分代码省略.........